Used Laboratory Equipment

CLARITY SOFTWARE PDA MODULE ADD ON - DATA APEX - Used Laboratory Equipment and New Lab Instrument Supplies, UK

Call us: 01204 848 430

You are here:

CLARITY SOFTWARE PDA MODULE ADD ON - DATA APEX

PDF  | Print |  E-mail

CLARITY SOFTWARE PDA MODULE ADD ON - DATA APEX CLARITY SOFTWARE PDA MODULE

The Clarity Software PDA module is a tool that is used for processing data that has been acquired from selected photo diode array detectors. Spectral data together with chromatograms add a third dimension to analytical data analysis.

The Clarity Software PDA module expands the capability of Clarity Chromatography software providing interactive spectral analysis, peak purity analysis and compound identification that is based on spectral library search. PDA data can be displayed in a set of optional views including 3D rendering.

The Clarity SoftwarevPDA module is optional addon software for the Clarity Chromatography Station (from version 2.4) and Clarity EVAL. The Clarity Chromatography Station is designed to acquire and evaluate data from up to four chromatographs at a time (multi-detector measurement support). Any Instrument within Clarity station can use the PDA module.

All data is saved in a single file; a chromatogram at any wavelength or only spectrum can simply be recalled after an analysis for review. UV-VIS spectra, acquired using a PDA detector, may be interactively selected from a chromatogram signal for visual inspection and comparison. The spectra may also be used for peak purity determinations, wavelength optimizations and component identification through spectral libraries.

Specification:CLARITY SOFTWARE PDA MODULE

Clarity Software PDA View: The PDA view is customisable; up to four views can be displayed at one time (Any combination from the following views may be selected: Isoplot, Chromatogram, Spectra, 3D Display, Peak Purity, Peak Purity Spectra, Library Spectra Search and Spectral Library). The user can easily extract chromatographic signals from PDA data to determine the optimal detection wavelength for each peak.

Clarity Software PDA Method: The Clarity PDA method includes an option for Spectral Library Search and Peak Purity analysis.

Spectral Library: The Clarity software compares the peak spectra with the spectra of an unlimited number of spectral libraries. Spectra stored in a Spectral Library include retention times and analysis parameters (optional). The Spectral Library Search can perform automatic identification of integrated/calibrated components (peaks). The library search may be constrained by the RT Window and by Wavelength Range. Either the Least Square, the Weighted Least Square or the Correlation Method is used for calculating library search matches. A Background Correction option is also available.

Peak Purity: This analysis helps to discover hidden impurities. Peak Purity test is applied to all integrated/calibrated peaks in the active signal. Purity is calculated from 5 or all spectra within the peak. The similarity curve is displayed in the PDA window. Peak Purity analysis can be optimised by setting custom preferences relating to the purity threshold, wavelength restriction and absorbance threshold.

Import/Export Data: Spectral data can be imported/exported in ASCII text formats to or from the Clarity software.

Reports: Users can easily include PDA options such as data 3D Display, Isoplot, Spectra, Peak Purity and Library Search results in reports using the intuitive Report Setup Dialog.

Instrument Control: A control module for Agilent 1100 DAD is available. Additional control modules for other PDA detectors can be developed upon request.

 