Used Laboratory Equipment

LIASYS - Fully automated random access benchtop Analyzer - Assel - Used Laboratory Equipment and New Lab Instrument Supplies, UK

Call us: 01204 848 430

You are here:

LIASYS - Fully automated random access benchtop Analyzer - Assel

PDF  | Print |  E-mail

Fully automated random access benchtop Analyzer - AsselFully automated random access benchtop Analyzer

 

LIASYS is a fully automated random access, continuos loading, benchtop Clinical Chemistry and Immunoturbidimetric analyzer. The instrument features primary tubes sampling, capacitive liquid level sensing, positive identification of the samples (optional), module for ISE (optional). It is therefore suitable to perform the daily routine of low/medium - sized laboratoriesin a fully safe and reliable way, providing flexible and cost effecive solutions. The LIASYS multitasking and user friendly software, running under Windows 98, allows the connection of two different analytical modules to the same computer to enhance the system potentialities and performances.

TECHNICAL FEATURES

 

DESCRIPTION:

 

Fully automated - random access, continuous loading, benchtop analyzer for Clinical Chemistry and Immunoturbidimetric assays.

 

ASSAY TIPE:

 

End Point, Kinetic, Differential, Bichromatic and Immunoturbidimetric assays; up to 100 different test, all user programmable.

 

TEST ENTRY MODE: selective - profiles- batch - STAT

 

ON LINE REAGENTS:

 

2 racks cooled by Peltier cells are provided to contain up to 20 reagents - 2 racks for 13 reagent positions at room temperature, 14 standards and controls
- 2 more positions for diluents.

 

SAMPLES LOADING: 5 racks for continuous loading of samples split in:

  • 64 positions for routine samples
  • 14 positions for STAT samples

PRIMARY TUBES SIZE: max Æ 16 x 100

 

STAT SAMPLES PROCESSING: allowed at any time

 

THROUGHPUT: 200 test/hour without ISE

 

WORKING TEMPERATURE: 37°C

 

READING SYSTEM: direct photometry

 

MINIMUM REACTION VOLUME: 300 ml

 

OPTICS SYSTEM

  • PHOTOMETER: multi - wavelenght optics
  • WAVELENGHT: 8 narrow band interferential filters
  • LIGHT SOURCE: halogen lamp 6V/10W
  • OPTICAL PATH: 10 mm
  • LINEAR RANGE: 0.001 - 2.500 Abs
  • RESOLUTION: 0,001 Abs

SAMPLING ARM: one mechanic arm provides all sampling operation with: · capacitive liquid level sensing · reagent pre-warming at 37°C · automatic probe washing

 

DILUTER: dilutions are made without using traditional syringes with the following specifications:

  • sample volume: 20 µl - 50 µl (1 µl incr.)
  • reagent volume: 25 µl - 400 µl (1 µl incr.)

REACTION PLATE: circular 60 semi-disposable cuvettes tray with following carachteristics:

  • automatic cuvette washing
  • cuvettes Q.C. continuously computer controlled
  • incubation temperature: 37°C

COMPUTER & SOFTWARE-

  • COMPUTER TYPE: IBM compatible
  • MEMORY: 32 Mb
  • FD: 1.44 Mb · HD: 3.2 Gb
  • MONITOR: colour SVGA 15" low radiation
  • REPORTER PRINTER: 80 columns impact graphic
  • INTERFACE: bidirectional RS232C
  • SOFTWARE: multitasking WINDOWS 98

OPTIONS

  • ISE module
  • Positive barcode reader
  • Barcode printer

INSTALLATION REQUIREMENTS

  • POWER SUPPLY: 90 - 250 Vac; 47/63 Hz; 300W
  • WATER CONSUMPTION: approx 3 ml/test

DIMENSION, WEIGHT & ENVIRONMENT

  • ANALYTICAL MODULE: Height: 42 cm; Depht: 65 cm; Lenght: 100 cm; Weight 65 Kg AMBIENT TEMPERATURE: 18°C - 30°C
  • RELATIVE HUMIDITY: 20% - 85%
 