On line Gas Analyzer - k2001ns New Series Trace Nitrogen in Argon or On Line Gas AnalyzerHelium

 

Contrôle Analytique, a leading manufacturer of trace gas analyzers since 1995, is now offering the K2001 new series with its patent pending plasma emission detector. The K2001 new series offers the legendary reliability of its predecessor plus a state of the art connectivity and an upgraded chemical interference free design.

The K2001, designed for a continuous measurement of nitrogen in argon, helium or both, is known for its unique stability over time that assures a nice repeatability for customers in need of a precise nitrogen quantification. These satisfied customers are located all around the world in different kinds of industries such as Air Separation Plants.

The K2001 new series is now up to date with a new microprocessor platform that allows a remote control of the analyzer through a local network or even through Internet. This connectivity feature allows you to control it from anywhere you want and to have Contrôle Analytique technicians assist you in the analyser's start-up or in any phase of diagnostic, without even being on site.

Contrôle Analytique is also proud to offer an exclusive autocalibration features that eliminates the need for the user to toggle calibration gases, avoiding time wastes and gas contamination (patented).

Features
  • Interference free
  • Guaranteed 10 ppb resolution / 2 ppb noise
  • Measuring system based on spectral emission of plasma maintained in a quartz chamber, controlled by a microprocessor system
  • Easy to operate
  • Solid state mass flow controller
  • Low sample flow (works with as low as 20 sccm)
  • Low power (35 watts)
  • No valves, columns, or carrier gas
  • No radioactive source
  • No moving parts, all solid state
  • No sensor or detector to replace
  • Maintenance free
  • Real Time Clock
  • Alarm Historic
  • CE conform and CSA on demand
  • Internet /Web enable
Applications
  • Air separation plants
  • Argon purification plants
  • Cryogenic truck loading station
  • Speciality gas laboratories
  • Process control
  • Steel industries
  • Chemical plants
  • Gas management system
  • Welding gas control
  • Helium liquification plants
Specifications
Dimensions

High (mm) Deep (mm) Wide (mm)
Standard rack mount 3U 5.25" (133.3) 18" (457.0) 19" (482.6)
Detector Type Electromagnetically induced plasma cell (dual beam)
Detector Material Quartz, single element, vacuum tight to 10 PSI (69kPA)/New Patent Pending Detector
Ranges
  • 0-1 ppm, resolution of 10 ppb, user configurable up to 10000 ppm
  • 0-10 ppm, resolution of 0.1 ppm, user configurable up to 10000 ppm
  • 0-100 ppm, resolution of 1 ppm, user configurable up to 10000 ppm
Accuracy Better than ± 1% full scale
Sample Flow Requirements 20 to 150 sccm
Gas Connections Sample: 1/8" Swagelok or 1/8" VCR tube fitting
Vent 1/8" swagelock tube fitting
Recommended Calibration Gas Zero: GP-200 purified gas (Getter)
Flow Accuracy 0 to 200 sccm ± 1% full scale
Recommended Maximum
Operating Pressure
20 PSIG (138 kPAG)
Recommended Minimum
Operating Pressure
4 PSIG (28 kPAG)
Operating Temperature 10 °C to 35 °C
Supply 115 VAC, 50 – 60 Hz or 220 VAC, 50 – 60 Hz
Power Consumption 35 watts (maximum)
Response Time 20 seconds for 90% of a step change at 75 sccm (typically 10 seconds)
Drift ± 1% over 24 hours (typically ± .5% over long term)
Weight 15 kg (33lbs)
Connectors
  • RJ-45 for Ethernet 10Base-T
  • DB-9 for RS-232 communication
  • Curtis GTFX-10 screw terminal block for I/Os
  • Corcom 5EFM4S power entry module
Standard Features
  • Microprocessor controlled with 4x40 LCD display and a 4x4 keypad
  • Self diagnostic system software
  • Manual or auto-ranging (user selectable)
  • Internet Web Enable
  • Real Time Clock
  • Alarm Historic
  • CE conform and CSA on demand
  • Digital ouputs for remote monitoring:
    (all dry relay contacts, SPST 1 amp 250 Vac resistive load)
    • System status (1 output)
    • Range in use (3 output)
    • 2 alarm outputs (user programmable set point)
Options
  • Auto-zero and auto-span calibration at user defined intervals
  • 4-20 mA isolated output:
    • 600 ohms at 20 mA dc
    • Fuse protection
    • Transient protected
    • Isolated from power ground and analyzer's electrical circuit
    • Programmable for over- and under-drive in case of a system alarm
    • Manual software control for calibration, monitoring and troubleshooting
  • Isolation: 750 VDC or AC peak maximum
  • Serial port: RS-232/422/485/modem
  • Custom option: contact factory
 