On Line Hydrocarbon Analyzer-k1000
|
PDF
|
| Print |
|
E-mail
|
|
On Line Hydrocarbon Analyzer - k1000
Contrôle Analytique, a leading manufacturer of trace gas analyzers since 1995, is now offering the K1000 hydrocarbon analyzer. The K1000 is now up to date with a new microprocessor platform that enables it to be remotely controlled from inside any area network and even Internet. This Internet connection is a very interesting feature. K1000 can be remotely controlled from everywhere by operators without being on site and allows Contrôle Analytique technicians to support those operators when they are in troubleshooting.
- Guaranteed 10 ppb resolution
- Easy to operate
- Easy to operate
- Alarm Historic
- Internet / Web Enable
|
- Graphic display
- All flow electronically regulated
- Temperature regulated
- Could be use with a methanizer for CO and CO2 measurement
|
- Air separation plants
- Semiconductor industries
- Speciality gas laboratory
|
- Environment
- Hydrocarbon measurement in hydrogen
- EPA method 25 capability
|
| |
High (mm) |
Deep (mm) |
Wide (mm) |
| Standard rack mount 3U |
5.25" (133.3) |
18" (457.0) |
19" (482.6) |
| Detector type |
High performance Low Noise Flame ionization detector sub system |
| Range |
0-1, 0-10, 0-100 ppm range (other ranges possible) |
| Sensitivity |
<= 10 ppb, less than 1% full scale (with temperature drift and pressure compensation) |
| Gas flow |
Recommended Flow:
- H2: 20 - 25 cc/min
- Air: 150 - 250 cc/min
- Sample: 10 - 40 cc/min at 5 to 40 PSIG (34 to 276 kPa)
NOTE: AIR AND H2 MUST BE HYDROCARBON FREE. THE USE OF HYDROCARBON TRAP ON CYLINDER IS RECOMMENDED. ALL FLOWS ARE DISPLAYED ON LCD SCREEN AND CONTROLLED BY HIGH PRECISION ELECTRONIC PRESSURE REGULATORS. H2 AND AIR REQUIRE THAT INLET PRESSURE IS 80 TO 100 PSIG (552 TO 690 KPA) |
| Detector temperature |
150 °C |
| Operating temperature |
10 to 35 °C |
| Gas connection |
1/8" compression fittings and 1/4", compression fittings for detector vent
NOTE: WATER COMES OUT OF DETECTOR VENT. A DOWN SLOPE LINE MUST BE USED TO AVOID WATER ACCUMULATION. KEEP IT AT AMBIENT TEMPERATURE TO AVOID FREEZING. |
| Supply |
115 Vac/50-60 Hz or 220 Vac 50-60 Hz (on order) |
| Fuse |
Two fuses 2 amp / 250 V, Little fuse «F», fast acting miniatures fuses, 5 x 20 mm
NOTE: FOR VOLTAGE SPIKES OVER 2 KV ON POWER INLET THE FUSES WILL OPEN DUE TO VARISTOR PROTECTIONS INSIDE THE UNIT. |
| Power consumption |
Maximum 125 watts |
| Weight |
15 kg (33 lbs) |
| Standard features |
- Real Time Clock
- Fully microprocessor-controlled, with graphic color display
- Self-diagnostic system software
- Fully remote controllable through Ethernet/Internet.
- Alarm Historic
- Manual or autoranging (user selectable)
- Digital ouputs for remote monitoring:
(all dry relay contacts, SPST 1 amp 250 Vac resistive load)
- System status (1 output)
- Range in use (3 output)
- 2 alarm outputs (user programmable set point)
|
| Option |
- Built-in auto-calibration (include valves)
- 4-20 mA isolated output
- Maximum load impedance: 600 ohms at 20 mA DC
- Isolation: 750 VDC or AC peak maximum
- Serial port: RS-232 / 422 / 485 / modem
- Custom option: contact factory
|
|