Plate Reader - Diareader - Dialab ELISA MICROPLATE READER (DIAREADER) MODELS: DIAREADER ELX800G: Measuring range from 400 nm to 750 nm For reading 24, 48 or 96-well plates DIAREADER ELX800 UV: Extended measuring range from 340 nm to 750 nm For reading 24, 48 or 96-well plates DIAREADER ELX800NB: Measuring range from 400 nm to 750 nm For reading 384-well microplates as well as 24, 48 and 96-well plates TECHNICAL SPECIFICATIONS - Microplates: 24, 48, 96 - well microplates(model ELX800NB also 384 wells)
- Speed: 30 seconds for 96 wells
- Wavelength range: 400-750nm (model ELX800UV 340 nm - 750 nm )
- Filters: 5-cavity wheel,
- Absorbance range: 0.000-3000 Abs
- Resolution: 0.001 Abs
- Accuracy: +/-1% +/- 0.0010 Abs from 0 to 2.000 Abs
- Linearity: +/-1% from 0 to 2.000 Abs, +/-3% from 2.000 to 3.000 Abs
- Repeatability: +/- 0,5% +/- 0.005 Abs from 0 to 2.000 Abs
- Software Interfacing: KC 3TM for WindowsTM, KC IITM, KC-JrTM, MacintoshTM Delta-SoftTM, Capture DataTM for ExcelTM
On-board data analysis: Simply to use and program via the reader "soft-keys" and 2x24 LCD display, the DIAREADER's on-board data reduction capabilities surpasses that of most PC software packages: - Extensive curve fitting: linear, cubic, quadratic, 4-P, log-logit, cubic-spline, point to point
- Auto or manual plate layout for blanks, controls, standards and samples
- Control and assay validation
- Transformation and Formula Applications
- Cut-off and Call Criteria
- Up to 75 assay programmed into DIAREADER's memory which can be recalled instantly
- Up to 8 microplates test results saved in the readers's memory
- Panel assay capable
- Store 25 standard curves
OPTIONS: Universal test plate - N IST traceable
- Checks alignment, accuracy, reproducibility and linearity
- Confirm PowerWave xTM and QuantTM wavelength accuracy with 7-filter test plate
- Use with KCjuniorTM for complete GLP compliance
KCjuniorTM: Basic data analysis and curve fitting - 32bit WindowsTM application controls all DIALAB Absorbance Readers
- Intuitive and flexible protocol creation
- Concise data analysis and reporting
- Flexible curve fitting options
- Contex-sensitive help screens
- Spectral scanning and analysis
- Pathlength correction with optional Bio-CellTM
Printer: HPTM or EPSONTM compatible printer ELECTRICAL DATA: Power: External, 24 volt power supply running from 100-240 VAC, +/- 10%, 50-60 Hz External outputs: RS 232 bidirectional serial interface, Parallel Centronics printer interface Display: 2x24 LCD display Regulatory: CE Mark, ETL listed, conforms to UL 3101-1, conforms to IEC 1010-1, certified to CAN/CSA C22.2 No. 1010.1 PHYSICAL DATA Dimensions: 41,9x38,1x17,8 cm (DxWxH) Weight: 8,4 kg |