Endolaser 476 - Enraf Nonius
|
PDF
|
| Print |
|
E-mail
|
|
Enraf Nonius Endolaser 476
The endolaser offers two types of laser therapy: continuous and pulsed. Continuous laser therapy is generally used for the treatment of semi-acute and chronic disorders, while pulsed therapy is mainly used for the treatment of very acute conditions. the maximum output is determined by the probe connected, and the power can be set in four steps(25,50,75 and 100)
Specifications
-
Laser type: solid state (ga-as-al)
-
Class: 3B(according to iec 825)
-
Laser wavelength: Max. 100mW (depending on probe). Adjustable to four levels 25% 50% 75% and 100% continuous/pulsed
-
Pulse repetition frequency: 300HZ (50% duty cycle)
-
Treatment time: max 99 min 59 seconds, depending on dose
-
Dose: max 9.99 joules
-
Warning signals: laser active (lamp/buzzer), operation error(buzzer), battery charge(lamp)
-
Tolerances laser power:10% of power selected
-
Laser test light meter:10% between 5 and 100mW
-
Probe
-
Laser wavelength:780nm
-
Power:10mW
-
Beam diameter: 4mm+ 1mm at10mm from probe
-
Angle of divergence 2.5
-
Target light wavelength:670nm
- Power:1mW
|
 |
|