Carbolite PF60 (200) Oven
|
PDF
|
| Print |
|
E-mail
|
|
Carbolite PF60 (200) Oven
Laboratory Oven
Bench Top Ovens |
|
This modern range of ovens provides a combination of excellent performance and reliability.
Increased power and low thermal mass encased fibre insulation ensure both fast heat up times and reduced recovery times. Reduced holding power once at set temperature, together with the insulation, makes the range economical and outer case temperatures have been significantly reduced.
Both gravity and forced air circulation models are available with a wide choice of control options allowing the most critical performance criteria to be met. Where processes involve the liberation of flammable vapours, a stoving and curing option is available. Also, where processes involve large amounts of water, a moisture extraction option is available.
| |
OVENS WITHOUT FANS |
OVENS WITH FANS |
| Model |
PN30 |
PN60 |
PN120 |
PN200 |
PF30 |
PF60 |
PF120 |
PF200 |
| Max Temp (°C) |
300 |
300 |
300 |
300 |
300 |
300 |
300 |
300 |
Chamber Dimensions(mm)
height
width
depth |
250
330
320 |
350
392
420 |
450
492
520 |
700
592
520 |
300
292
320 |
400
392
420 |
500
492
520 |
750
592
520 |
External Dimensions(mm)
height
width
depth |
470
665
470 |
570
765
570 |
670
865
670 |
920
965
670 |
470
665
470 |
570
765
570 |
670
865
670 |
920
965
670 |
| Chamber Capacity (liters) |
27 |
58 |
115 |
215 |
28 |
66 |
128 |
230 |
| Weight (Kg) |
30 |
45 |
60 |
75 |
30 |
45 |
60 |
75 |
Shelves
(number supplied)
(max possible)
(max dist load/shelf kg)
(max load kg) |
2
3
10
20 |
2
5
10
30 |
2
9
10
40 |
2
15
10
50 |
2
3
10
20 |
2
5
10
30 |
2
9
10
40 |
2
15
10
50 |
| PERFORMANCE |
|
|
|
|
|
|
|
|
Power Rating
at 240 V (watts) |
750 |
1000 |
1500 |
2250 |
1000 |
1500 |
2000 |
2700 |
Holding Power*
at max temp (watts) |
300 |
480 |
720 |
1160 |
350 |
600 |
800 |
1250 |
Temperature Uniformity*
(at max temp as a %) |
+/-2.3 |
+/-2.3 |
+/-2.7 |
+/-3.5 |
+/-1.0 |
+/-1.0 |
+/-1.0 |
+/-1.0 |
Temperature Stability
on/off control (°C) |
+/-1.5 |
+/-1.5 |
+/-1.5 |
+/-1.5 |
+/-1.0 |
+/-1.0 |
+/-1.0 |
+/-1.0 |
Heat Up Times* (mins) 240V
100°C
200°C
300°C |
12
26
52 |
12
26
52 |
12
26
52 |
14
29
58 |
4.5
12
25 |
4.5
12
25 |
4.5
12
25 |
5.5
14
30 |
Recovery Times* (mins)
door open 60secs 240V
100°C
200°C
300°C |
2.5
5
8.5 |
2.5
5
8.5 |
2.5
5
8.5 |
3
6
10 |
1
2.5
4 |
1
2.5
4 |
1
2.5
4 |
1.5
3
5 |
Air Exchanges vol (l/h) @100°C
(a)
(b) |
N/A
N/A |
N/A
N/A |
N/A
N/A |
N/A
N/A |
1400
10,000 |
1400
10,000 |
1400
10,000 |
1400
10,000 |
Air Exchanges - exchanges/hour
(a)
(b) |
N/A
N/A |
N/A
N/A |
N/A
N/A |
N/A
N/A |
50
360 |
21
153 |
11
79 |
6
44 |
Note: A uniformity of +/-1% = +/-1°C at 100°C.
* With vents closed
(a) with standard fan
(b) with optional exhaust fan |
|
|