Used Laboratory Equipment

591 Star Flow Volumetric Pump - Used Laboratory Equipment and New Lab Instrument Supplies, UK

Call us: 01204 848 430

You are here:

591 Star Flow Volumetric Pump

PDF  | Print |  E-mail

 Ivac 591 Star Flow Volumetric Pump

 

Similar in appearance and operation to the IVAC 580 series, these devices have increased accuracy to ± 5 %. Both have linear peristaltic operation, integral pole clamp, secondary infusion capability, computer interface, dose limit, and are housed in a lightweight impact resistant plastic insulated housing. The 590 requires a drop sensor, while the 599 runs without.

 

Rate parameters:

1-999 ml/hr in 1 ml increments

Volume, Dose Limits:

Off-9999 ml for primary, Off-999 for secondary

Pressure Limits:

Occludes at < 25 PSI.

Battery capacity:

Fully-charged lasts approx 6 hours at 125 ml/hr

Battery recharge:

Discharged to full charge in approx. 24 hours.

KVO Rate:

5 ml/hr or selected rate if lower.

Alarms:

Flo, Door, Hold, Bat., Batt., Set Out, Err X, Occl., Air

Accuracy:

± 5 %

Flow Sensor:

Model 192 for 590, no sensor required for 599

 IVAC590.jpg

 

 
 